1
Negotiate Terms with Seller
Discuss and agree upon the equipment's condition, configuration, and pricing with the seller before proceeding.
Brand | DRAGER |
Condition | Certified Refurbished |
Model | Fabius GS |
Warranty | Not specified |
The Drager Fabius GS anesthesia machine is a popular choice for practitioners who demand advanced ventilation technology combined with reliable and proven breathing system components. The Fabius GS anesthesia workstation opens new frontiers in ventilation performance while maintaining well-established functionality.
High-performance ventilation with a versatility that’s more flexible and economical to use than traditional gas-driven bellows ventilators
Full-color screen
Compact, convenient breathing system that’s ergonomically designed for optimal positioning
ClinicalVision – monitoring solution that allows the caregiver to view patient monitoring and clinical data
simultaneously
Customized design for flexibility and ease of use
Ready for IT-integration
Low-flow ventilation
Electronic vertical flow controls and electronic fresh gas flow indicators
Intelligent cable management for reduced clutter and easier patient trans
Easy integration into an existing hospital information system
Central braking
Continuous monitoring during transport and automatic reconfiguration of site-specific parameters with Dräger’s innovative Infinity Docking Stations (IDS).
Drager Fabius GS Specifications
Dimensions
Height (with COSY): 51.5″ (130.8 cm)
Width (with COSY): 35″ (89 cm)
Depth (with COSY): 33.1″ (84 cm)
Weight: 224 lbs (101.6 kg)
Power Supply
100 – 240 VAC, 50/60 Hz, 2.3 A max.
Battery (supports ventilator and monitor) > 45 min.
Rechargeable Batteries
Rating: 24V; 3.5 Ah
Type: Sealed, gelled lead-acid
Recharge Time: ≤ 16 hours on the mains or full operation time
Operating time with fully charged batteries: 45 minutes, minimum
Ventilator Input Ranges
PMAX Pressure Limiting: 15 – 70 cmH2O (1 cmH2O resolution)
VT Tidal Volume: 20 – 1400 mL (10 mL resolution)
VT (SIMV/PS) Tidal Volume: 20 – 1100 mL (10 mL resolution)
f Breathing Frequency: 4 – 60 bpm (1 bpm resolution)
Ti/Te Inspiration/Expiration ration: 4:1 to 1:4
Tip/Ti Inspiration pause: 0% – 50% (1% resolution)
PEEP End-expiratory Pressure: 0 – 20 cmH2O (1 cmH2O resolution)
Pinup Inspiratory Pressure: 5 – 65 cmH2O (1 cmH2O resolution)
Insp Flow Inspiratory Flow: 10 – 75 L/min (1 L/min resolution in Pressure Control Mode; 10 – 85 L/min (1 L/min resolution) in PS and SIMV/PS modes
Δ PPS (Pressure Support) Support Pressure: 3 – 20 cmH2O (1 cmH2O resolution)
Δ PPS (SIMV/PS) Support Pressure : 3 – 20 cmH2O, OFF (1 cmH2O resolution)
Freq Min Apnea Ventialtion minimum frequency: 2 – 20 bpm (1 bpm resolution) and OFF”
Trigger: Trigger Level: 2 – 15 L/min (1 L/min resolution)
Tinsp SIMV Inspiratory Time: 0.3 – 4.0 sec
Pressure Support Ventilation Mode
Endotracheal Tube Size: 4.5 mm – 8 mm
Patient Lung Compliance: 10 ml/cmH2O – 100 ml/chH2O
Unassisted Patient Tidal Volume: 50 ml – 1000 ml
Patient Breath Rate (BPM): 10 – 35
Delivery Accuracy
Pmax Pressure Limiting: ±5 cmH2O of setting
VT Tidal Volume: ± 5% of setting or 20 mL, whichever is greater
f Breathing Frequency: ±1 bpm of setting
Ti/Te Inspiration/Expiration Ratio: ±5% of setting
Tip/Ti Inspiration Pause: ±25% of setting
PEEP End-expiratory Pressure: ±2 cmH2O or ±20% of setting, whichever is greater
Safety Reliefs
High-Pressure Safety Relief Valve: 75 ±5 cmH2O
Negative Pressure Safety Relief Valve (Ambient Air Inlet Valve): -7.5 to -9 cmH2O
System Compliance Compensation Measurement: 0.2 to .0 ml/cmH2O: ±0.2 ml/cmH2O or ±10% of actual compliance, whichever is greater
Anesthesia Gas Supply Module
Fresh Gas Flow Indicators: O2, N2O, Air: Range and accuracy: 0.0 to 12.0 L/min ±10% of reading or 0.12 L/min (into an ambient atmosphere of 14.7 psi (101.3 kPa) at 20°C). Resolution: 0.1 L/min
Fresh Gas Flow Stability: O2 and N2O: ±10% of setting with pipeline pressures between 45 – 65 psi; Air ±10% of setting with pipeline pressures between 50 – 55 psi; Air flow rate will vary proportionally with supply pressures outside 50 – 55 psi
Total Fresh Gas Flowmeter: Range and accuracy: 0 to 10 L/min ±10% of full scale at STP, calibrated with 50% O2 / 50% N2O gas mixture; 0 to 10 L/min ±15% of full scale at STOP for all other gas mixtures; Resolution: 0.5 L/min from 0.5 – 2 L/min; 1.0 L/min from 2 – 10 L/min
O2 Flush (bypass): at 55 psi (3.8 bar) max. 50 L/min; at 50 psi (3.4 bar) min. 35 L/min
Common Gas Outlet Pressure Limit: 13 psi (0.9 bar)
Anesthetic Agent Vaporizer Interface
The Drager Vapor quick-change plug-in system for up to two anesthetic agent vaporizers. The connections are automatically closed and sealed when the vaporizer is removed.
Drager Halothane Vapor
Drager Enflurane Vapor
Drager Isoflurane Vapor
Drager Sevoflurane Vapor
Datex-Ohmeda Devapor/D-Tech for Desflurane
Drager D-Vapor
Brand | DRAGER |
Condition | Certified Refurbished |
Model | Fabius GS |
Warranty | Not specified |
The Drager Fabius GS anesthesia machine is a popular choice for practitioners who demand advanced ventilation technology combined with reliable and proven breathing system components. The Fabius GS anesthesia workstation opens new frontiers in ventilation performance while maintaining well-established functionality.
High-performance ventilation with a versatility that’s more flexible and economical to use than traditional gas-driven bellows ventilators
Full-color screen
Compact, convenient breathing system that’s ergonomically designed for optimal positioning
ClinicalVision – monitoring solution that allows the caregiver to view patient monitoring and clinical data
simultaneously
Customized design for flexibility and ease of use
Ready for IT-integration
Low-flow ventilation
Electronic vertical flow controls and electronic fresh gas flow indicators
Intelligent cable management for reduced clutter and easier patient trans
Easy integration into an existing hospital information system
Central braking
Continuous monitoring during transport and automatic reconfiguration of site-specific parameters with Dräger’s innovative Infinity Docking Stations (IDS).
Drager Fabius GS Specifications
Dimensions
Height (with COSY): 51.5″ (130.8 cm)
Width (with COSY): 35″ (89 cm)
Depth (with COSY): 33.1″ (84 cm)
Weight: 224 lbs (101.6 kg)
Power Supply
100 – 240 VAC, 50/60 Hz, 2.3 A max.
Battery (supports ventilator and monitor) > 45 min.
Rechargeable Batteries
Rating: 24V; 3.5 Ah
Type: Sealed, gelled lead-acid
Recharge Time: ≤ 16 hours on the mains or full operation time
Operating time with fully charged batteries: 45 minutes, minimum
Ventilator Input Ranges
PMAX Pressure Limiting: 15 – 70 cmH2O (1 cmH2O resolution)
VT Tidal Volume: 20 – 1400 mL (10 mL resolution)
VT (SIMV/PS) Tidal Volume: 20 – 1100 mL (10 mL resolution)
f Breathing Frequency: 4 – 60 bpm (1 bpm resolution)
Ti/Te Inspiration/Expiration ration: 4:1 to 1:4
Tip/Ti Inspiration pause: 0% – 50% (1% resolution)
PEEP End-expiratory Pressure: 0 – 20 cmH2O (1 cmH2O resolution)
Pinup Inspiratory Pressure: 5 – 65 cmH2O (1 cmH2O resolution)
Insp Flow Inspiratory Flow: 10 – 75 L/min (1 L/min resolution in Pressure Control Mode; 10 – 85 L/min (1 L/min resolution) in PS and SIMV/PS modes
Δ PPS (Pressure Support) Support Pressure: 3 – 20 cmH2O (1 cmH2O resolution)
Δ PPS (SIMV/PS) Support Pressure : 3 – 20 cmH2O, OFF (1 cmH2O resolution)
Freq Min Apnea Ventialtion minimum frequency: 2 – 20 bpm (1 bpm resolution) and OFF”
Trigger: Trigger Level: 2 – 15 L/min (1 L/min resolution)
Tinsp SIMV Inspiratory Time: 0.3 – 4.0 sec
Pressure Support Ventilation Mode
Endotracheal Tube Size: 4.5 mm – 8 mm
Patient Lung Compliance: 10 ml/cmH2O – 100 ml/chH2O
Unassisted Patient Tidal Volume: 50 ml – 1000 ml
Patient Breath Rate (BPM): 10 – 35
Delivery Accuracy
Pmax Pressure Limiting: ±5 cmH2O of setting
VT Tidal Volume: ± 5% of setting or 20 mL, whichever is greater
f Breathing Frequency: ±1 bpm of setting
Ti/Te Inspiration/Expiration Ratio: ±5% of setting
Tip/Ti Inspiration Pause: ±25% of setting
PEEP End-expiratory Pressure: ±2 cmH2O or ±20% of setting, whichever is greater
Safety Reliefs
High-Pressure Safety Relief Valve: 75 ±5 cmH2O
Negative Pressure Safety Relief Valve (Ambient Air Inlet Valve): -7.5 to -9 cmH2O
System Compliance Compensation Measurement: 0.2 to .0 ml/cmH2O: ±0.2 ml/cmH2O or ±10% of actual compliance, whichever is greater
Anesthesia Gas Supply Module
Fresh Gas Flow Indicators: O2, N2O, Air: Range and accuracy: 0.0 to 12.0 L/min ±10% of reading or 0.12 L/min (into an ambient atmosphere of 14.7 psi (101.3 kPa) at 20°C). Resolution: 0.1 L/min
Fresh Gas Flow Stability: O2 and N2O: ±10% of setting with pipeline pressures between 45 – 65 psi; Air ±10% of setting with pipeline pressures between 50 – 55 psi; Air flow rate will vary proportionally with supply pressures outside 50 – 55 psi
Total Fresh Gas Flowmeter: Range and accuracy: 0 to 10 L/min ±10% of full scale at STP, calibrated with 50% O2 / 50% N2O gas mixture; 0 to 10 L/min ±15% of full scale at STOP for all other gas mixtures; Resolution: 0.5 L/min from 0.5 – 2 L/min; 1.0 L/min from 2 – 10 L/min
O2 Flush (bypass): at 55 psi (3.8 bar) max. 50 L/min; at 50 psi (3.4 bar) min. 35 L/min
Common Gas Outlet Pressure Limit: 13 psi (0.9 bar)
Anesthetic Agent Vaporizer Interface
The Drager Vapor quick-change plug-in system for up to two anesthetic agent vaporizers. The connections are automatically closed and sealed when the vaporizer is removed.
Drager Halothane Vapor
Drager Enflurane Vapor
Drager Isoflurane Vapor
Drager Sevoflurane Vapor
Datex-Ohmeda Devapor/D-Tech for Desflurane
Drager D-Vapor
Write us an email!
We will call you back soon!
Contact us