1
Negotiate Terms with Seller
Discuss and agree upon the equipment's condition, configuration, and pricing with the seller before proceeding.
Brand | NARANG |
Condition | New |
Model | DP4300 |
Warranty | 12 month(s) |
Features :
› Microprocessor integrated with temperature sensors provides high precise control.
› Multiple independent over-heating protections and automatic cut off.
› Visual and acoustic alarm for malfunction.
› Open system to accept standard IV tube, no need for special disposables.
› Quick warming up and easy operation.
Distinguish Features :
For massive infusions, simultaneous
warming of multiple infusions, with 5m.,
IV tube
Flow Rate-KVO : 6000ml/hr, 100ml/min
Temperature Setting : 37°C-42°C/
0.1 INCR.
Power Supply : A.c.240V±10%/ 50-60Hz
Power Consumation : 250 Watt
Peak Absorption : Max 250 Watt
Protection Class : CLASS I
Degree of Protection : TYPE B
Degree of Protection Liquids : IPX4
Fuse : T1.6A/250VAC
Mdd Classification : IIb
Overheat Protection : 43°C/45°C/48±3°C
Alarms Low-Hith Temp. : 36°C / 43°C
Warming Time : <2 minutes (5°C – 36°C)
<45seconds ( 20°C – 36°C)
Operating Mode : Continuous
Dimensions : 100x170x210 mm
Weight : KG. 1.3
Leadtime within 7-30 days after payment received. Depends on availability of the product in the warehouse.
Brand | NARANG |
Condition | New |
Model | DP4300 |
Warranty | 12 month(s) |
Features :
› Microprocessor integrated with temperature sensors provides high precise control.
› Multiple independent over-heating protections and automatic cut off.
› Visual and acoustic alarm for malfunction.
› Open system to accept standard IV tube, no need for special disposables.
› Quick warming up and easy operation.
Distinguish Features :
For massive infusions, simultaneous
warming of multiple infusions, with 5m.,
IV tube
Flow Rate-KVO : 6000ml/hr, 100ml/min
Temperature Setting : 37°C-42°C/
0.1 INCR.
Power Supply : A.c.240V±10%/ 50-60Hz
Power Consumation : 250 Watt
Peak Absorption : Max 250 Watt
Protection Class : CLASS I
Degree of Protection : TYPE B
Degree of Protection Liquids : IPX4
Fuse : T1.6A/250VAC
Mdd Classification : IIb
Overheat Protection : 43°C/45°C/48±3°C
Alarms Low-Hith Temp. : 36°C / 43°C
Warming Time : <2 minutes (5°C – 36°C)
<45seconds ( 20°C – 36°C)
Operating Mode : Continuous
Dimensions : 100x170x210 mm
Weight : KG. 1.3
Leadtime within 7-30 days after payment received. Depends on availability of the product in the warehouse.